| The reason to use SFA or FIA | | | | approvals. Since the EPA considers FIA equivalent to |
| The reason to automate is so that you can run more | | | | SFA, OI did not obtain EPA ATP letters. There are |
| samples. Running more samples is the same as | | | | also several FIA methods that are EPA approved |
| increasing your productivity. Having an instrument run | | | | without an SFA equivalent, for instance CN by |
| samples for you decreases your labor costs. Having | | | | OIA1677. |
| the instrument do all those boring mixings and | | | | In Flow Injection analysis the sample is injected by a |
| shakings increases your day to day precision. This | | | | valve into a carrier stream. The sample solution does |
| improves quality and really helps with that QA/QC | | | | not pass through peristaltic pump tubing prior to the |
| program. Whether you are a research facility, a | | | | valve. Mixing occurs as the sample travels through |
| municipal laboratory, or a commercial lab, you can | | | | tightly wound Teflon mixing coils. Unlike segmented |
| benefit from automation. | | | | flow that may contain multiple samples within the |
| Almost any method that can be done manually can | | | | tubing all at the same time, Flow injection injects a |
| be automated. Time consuming steps such as manual | | | | sample and detects it before the next sample is |
| titrations can be replaced using instrumentation. | | | | injected. |
| Instruments can do digestions, distillations, dilutions, | | | | Comparison of SFA and FIA |
| and filtrations all for you. More importantly, is that | | | | Flow Injection is basically a derivative of SFA without |
| these methods are always duplicated exactly | | | | air segmentation. Both SFA and FIA mix sample and |
| ensuring that quality control procedures are met. | | | | reagents in a continuously flowing stream of |
| Your most expensive cost is labor, and after that is | | | | reagents. The reaction that occurs, the peak shape, |
| supplies. A flow analyzer will allow you to do more | | | | the sensitivity, and so forth is all determined by how |
| samples in less time and increase the amount of time | | | | the tubing is configured, where the reagents are |
| you or your staff spends on more important things | | | | teed in, and the length of the tubing that the |
| like running other samples, reporting results, and doing | | | | reactions are happening in. Instrument manufacturers |
| that dreaded paper work. Automated equipment | | | | configure tubing "cartridges" according to published |
| uses less reagents and generates less waste than | | | | methodology, or develop methods based on |
| manual methods do. This is not about getting rid of | | | | R&D and published works. |
| staff. This is about freeing up time so that staff can | | | | An analyst can readily change and/or modify method |
| have time to prepare and run more samples. The | | | | performance simply by modifying any aspects of the |
| beauty of the auto analyzer is that it is running | | | | analytical cartridge. Since the segmentation of SFA |
| samples while you are doing something else. | | | | limits dispersion, SFA is a bit more forgiving than FIA |
| History of Flow Analysis | | | | in the length of tubing. For instance, the addition of |
| The original automated analyzer is the Auto Analyzer | | | | extra tubing on a FIA method can significantly change |
| invented by Leonard Skeggs in 1954. It took the | | | | peak shape and sensitivity, while additional tubing on |
| Technicon Corporation three years to perfect it and | | | | a SFA system does not really matter. For example, |
| develop a commercial product. The concept of | | | | at one of my previous laboratories we routed an |
| auto-analysis eventually branched out and has roots | | | | SFA method that had a non-working heater cartridge |
| in almost every automated device we see today. | | | | through several feet of tubing to the fecal coliform |
| Flow Injection Analysis was a way to get out of | | | | bath and then back to the detector. Adding several |
| buying a Technicon Auto-Analyzer because a person | | | | feet of tubing would have been detrimental to FIA, |
| with a peristaltic pump, some Teflon tubing, and a | | | | but was un noticeable in the SFA method. |
| flow through spectrometer cell could fabricate their | | | | Literature is deceiving in sample size as SFA methods |
| own device. In fact, the first FIA instruments were | | | | tend to use less sample than FIA. Throughput by FIA |
| held together with Lego blocks. The first commercial | | | | is not faster, and in fact it has been demonstrated |
| FIA instrument was introduced by Tecator, the | | | | that a well operated SFA system far exceeds FIA in |
| analytical branch of Perstorp. | | | | samples per hour. SFA mixes by an end over end |
| All continuous flow analyzers share a commonality of | | | | similar to putting sample and reagent in a vial and |
| parts with similar functions. All have an autosampler, | | | | inverting. FIA mixes by what Ruzika termed |
| pumps, reagent mixing manifolds comprised of tubing, | | | | "controlled dispersion" which is harder to explain than |
| a flow through detector, and a way to collect the | | | | end over end, and is a result of solution flowing |
| signal. The signal is always expressed as a peak with | | | | faster through the center than at its sides, and then |
| a maximum and a baseline, and the peak is either bell | | | | rapidly colliding with the walls of the mixing coils and |
| shaped, or rectangular. | | | | other flow interruptions placed in its path. SFA signals |
| Segmented Flow Analysis (SFA) | | | | do not have to be brought to a maximum, but |
| Segmented Flow analysis is the first auto chemistry | | | | theoretical SFA should always be and the resulting |
| and is what Skeggs initially demonstrated to the | | | | peak would be rectangular. Because of the "tunneling" |
| Technicon corporation. Segmented flow enjoys over | | | | that occurs in FIA flow, a FIA peak is Gaussian. |
| 50 years of success as a proven technology. | | | | It is hard to say exactly what the maximum time for |
| Segmented flow is the basis of multiple automated | | | | a SFA incubation is because the only limiting factor is |
| EPA and many other regulatory methods. | | | | the minimal carryover that occurs as solution |
| A typical segmented flow analyzer pumps the sample | | | | "transfers" from one segment to the next while |
| into the cartridge using peristaltic pump tubing. The | | | | traveling down the tube. Most FIA literature places 2 |
| sample merges with an air segmented carrier stream, | | | | minutes at the maximum for FIA. In fact, Ruzika and |
| reagents are added and mixing occurs by end over | | | | Hansen gave a 1 minute maximum for FIA reaction. |
| end mixing in coils as the sample is transported down | | | | Advantages of Flow Methods |
| the path of the tubing. Depending on the internal | | | | Many methods are written specifically for continuous |
| tubing volume and the amount of time the transport | | | | flow analysis. Some examples are ammonia by EPA |
| from injection to detection takes there could be | | | | 350.1 or Block digestion TKN by EPA 351.2. An |
| multiple samples inside the tubing at a given time. For | | | | example of a FIA method written specifically for FIA |
| instance, if the delay time is 10 minutes and samples | | | | is OIA 1677. These EPA methods specify automated |
| are injected every minute there will be 10 samples | | | | chemistry in the methodology itself. Other methods, |
| traveling within the tubing. | | | | such as the manual fluoride by ISE method, can be |
| Flow Injection Analysis (FIA) | | | | modified according to 40 CFR Part 136.6 and made |
| Flow Injection Analysis is considered as an acceptable | | | | into automatic methods. Continuous flow methods |
| alternative to equivalent segmented flow methods. In | | | | reduce manual labor and that reduction of manual |
| other words, though USEPA method 350.1 is a | | | | labor tends to make precision better and reduce |
| segmented flow method, the USEPA considers a FIA | | | | potential contamination. Flow methods also tend to |
| method equivalent. There are thousands of literature | | | | have lower MDL's than manual methods and a better |
| references to FIA methods, and multiple ATP | | | | chance of fewer QC failures. |